Instrument Resolution: Models and Calibration

A measured resonance dip is broadened by several largely separate effects: the instrument resolution (moderator pulse + flight-path + detector timing — a beamline property), Doppler broadening (sample temperature), the background / normalization (the measurement), and self-shielding / multiple scattering in optically-thick (black) resonances (the sample). On a single spectrum these trade off — too narrow a resolution looks like too hot a sample — so the instrument resolution must be characterized separately and then held fixed when fitting unknown samples.

This page covers (1) the three resolution models NEREIDS provides and (2) the calibrate → pin → fit procedure that determines the resolution from a known standard.

The three resolution models

All three are passed to forward_model and the fitters the same way; only the kernel source differs.

Gaussian (analytical)

Energy-dependent Gaussian from instrument timing parameters:

import numpy as np, nereids
hf = nereids.load_endf(72, 177)              # Hf-177
energies = np.linspace(2.0, 30.0, 2000)

T = nereids.forward_model(
    energies, [(hf, 5e-5)], temperature_k=300.0,
    flight_path_m=25.0, delta_t_us=1.0, delta_l_m=1e-3,
)

Tabulated UDR (Monte-Carlo file)

UDR = User-Defined Resolution — SAMMY’s term for a numerical resolution function supplied as a (time/energy, weight) table (SAMMY manual, “User-Defined Numerical Resolution Function”). In NEREIDS this is a measured/simulated asymmetric kernel, e.g. a VENUS FTS file:

udr = nereids.load_resolution("fts_bl10.txt", flight_path_m=25.0)
T = nereids.forward_model(energies, [(hf, 5e-5)], temperature_k=300.0, resolution=udr)

Ikeda–Carpenter (analytical moderator model)

A physically-grounded analytic moderator pulse — α(E) (fast rate), β (slow storage rate), R (storage fraction), with optional proton-burst and channel terms. Build parameter laws with EnergyLaw, then synthesize the kernel:

ic = nereids.IkedaCarpenter(
    flight_path_m=25.0, e_min_ev=0.5e-3, e_max_ev=1000.0,
    alpha=nereids.EnergyLaw.sqrt_e(0.30, 0.0),   # α(E) = 0.30·√E
    beta=0.10,
    r=nereids.EnergyLaw.exp_mev(25.0),           # R(E) → 0 in the eV regime
)
tab = ic.as_tabulated()                          # a TabulatedResolution
T = nereids.forward_model(energies, [(hf, 5e-5)], temperature_k=300.0, resolution=tab)

ic.kernel_at(energy_ev) returns the (tof_offsets_us, weights) kernel at one energy for inspection.

Performance note. An IkedaCarpenter synthesizes its kernel table once at construction and caches it (as_tabulated() just clones it). The cost to know about is that IC is not plan-cached when fitting the t0/L energy-scale at run time: unlike a loaded TabulatedResolution, its broadening plan is rebuilt each energy-scale evaluation, so an IC run-time fit over a large grid is slower than the tabulated path. Resolution calibration is once-per-experiment, so the calibrate → pin → fit workflow is unaffected; for production spatial maps, pin the calibrated kernel via as_tabulated() (which is plan-cached).

The calibrate → pin → fit procedure

Instrument resolution and flight-path geometry are beamline constants; density, temperature, background and normalization are per-measurement. So:

  1. Calibrate — measure a calibrant of known density and temperature; fit the resolution parameters with ρ, T fixed.
  2. Pin — keep the calibrated resolution; switch to the sample (same geometry).
  3. Fit — fit the sample ρ / T / both with the resolution pinned; re-fit background and normalization per measurement.

calibrate_resolution

cal = nereids.calibrate_resolution(
    energies, data, uncertainty,
    family="udr_corr",                  # "gaussian" | "udr_corr" | "ic"
    isotopes=[(hf, 5e-5)],              # KNOWN calibrant composition + density
    temperature_k=300.0,                # KNOWN calibrant temperature
    base_udr=udr,                       # required for family="udr_corr"
    restarts=2,
)
print(cal)                # ResolutionCalibration(family=udr_corr, chi2/dof=..., converged=..., n_free_params=2, bounds_hit=[])
print(cal.params())       # decoded fitted parameters
calibrated = cal.as_tabulated()         # pin this into the sample fit

The families calibrate different knobs:

familyfitsmeaning
gaussianΔt, ΔLanalytical Gaussian width
udr_corrs(E)=s0·(E/E_ref)^p on a base UDRtrust the MC shape, calibrate its width
icα(E)=e^c0·√E+e^c1, β, R (⊗ PSR triangle)full bounded analytic moderator shape

The ic family fits the complete bounded Ikeda–Carpenter shape: the prompt law α(E) = e^{c0}·√E + e^{c1} is positive at every energy by construction (the coefficients are exp-encoded — a calibration can no longer return an a1 < 0 that flips α negative below the fit window), and the storage rate β and mixing fraction R ∈ [0, 1] are free within physics bounds. The kernel is convolved with the SNS PSR (accumulator-ring) channel triangle — FWHM psr_fwhm_ns (default 350 ns, the value the VENUS FTS file header records as already folded into the tabulated kernel; 0 disables). Pass fit_psr=True to also fit the triangle FWHM as a 5th parameter. The PSR fold applies to ic only: tabulated/UDR kernels already carry it in the file and are never re-folded.

Every result echoes n_free_params and bounds_hit — a list of "name:lower" / "name:upper" strings for parameters pinned at a box bound. A pinned bound flags a degenerate direction: e.g. an eV-regime calibrant with no storage tail drives R → 0 ("r:lower"), and on that β↔R ridge the reported β carries no information.

Use .as_tabulated() for udr_corr / ic (a TabulatedResolution to pass as resolution=); use .gaussian_params()(delta_t_us, delta_l_m) for the Gaussian family. For ic, cal.params() returns the decoded {a0, a1, beta, r, psr_fwhm_us} (the raw theta is ln/box-encoded optimizer space).

Pin and fit the sample

fit = nereids.fit_spectrum_typed(
    sample_T, sample_unc, energies, [(hf, 1e-4)],
    temperature_k=300.0, fit_temperature=True,   # recover ρ and T
    resolution=calibrated,                        # PINNED
)
print(fit.densities, fit.temperature_k, fit.reduced_chi_squared)

Choosing a calibrant (important)

The fit absorbs every unmodeled broadening into the “resolution”, so a poor calibrant yields a contaminated, non-transferable result:

  • Use non-black resonances (T_min ≈ 0.2–0.8). Geometric thinness is not enough — at a strong resonance the cross-section is thousands of barns, so even a thin foil is optically thick at the peak and its self-shielding / multiple scattering would be soaked into the “resolution”.
  • Model the background / normalization during calibration (fit_background=True), then re-fit it per sample (it does not transfer).
  • Same geometry for calibrant and sample (flight path, sample-to-detector).

Guidance

  • Density / isotopic characterization is robust to the resolution-model choice; temperature is sensitive — calibrate carefully before trusting a fitted temperature.
  • The resolution width is well-determined; the resolution family is not, from a calibrant alone. The ic kernel anchors its mode at zero offset, so a right-skewed pulse’s centroid lags by ~1/α(E), shifting a broadened dip’s apparent energy. For the a1=0 prompt law that lag is ≈1/√E — the same basis as a flight-path (L_scale) error (leading-order for a1≠0) — so absolute dip position cannot distinguish an asymmetric kernel from a small L miscalibration. By default calibrate_resolution pins the energy scale (fit_t0=fit_l_scale=False): a pure shape/width fit on the already energy-calibrated grid. The cross-family χ² then discriminates on shape and position, which is honest only when (t0, L) are independently known; otherwise the position part is confounded with L.
  • To handle position honestly, fit the shared energy scale under a metrology prior. Set fit_t0=True / fit_l_scale=True (centered at t0_center_us / l_scale_center, with Gaussian priors t0_prior_us / l_scale_prior from the instrument’s flight-path / timing metrology). The fit reports position_t0_us, position_l_scale, and prior_penalty. Do not fit position with a flat prior — a free L_scale absorbs the asymmetric-kernel lag and corrupts the calibrated width. With a weak prior, family discrimination collapses toward the position-independent skew/tail evidence only (χ²/dof ≈ 1.1–1.3 in synthetic Hf-177 studies), so report it as a function of the prior strength rather than claiming strong discrimination.
  • A worked end-to-end example (build the models, calibrate, pin, fit) is in examples/notebooks/workflows/06_resolution_calibration.ipynb.